Carbon Monoxide Sensors

CO molecule
  • 3-series conforming standard size

    AWX CO-3 1000

    Standard 3-Series

    • Nominal Range 0 – 1000 ppm
    • Maximum Overload 2000 ppm
    • Expected Operation Life 3 years in air
    • Output Signal 120 +/- 20 nA/ppm
    • Resolution 1.0 ppm
    • Temperature Range - 30 °C to + 30 °C
    • Pressure Range Atmospheric +/-10 %
    • t90 Response Time < 30 sec
    • Relative Humidity Range 15 % to 90 % R.H. non-condensing
    • Typical Baseline Range (pure air, 20°C) – +/- 7 ppm
    • Maximum Zero Shift (+20°C to +40 °C) 2.0 ppm
    • Expected Long Term Output Drift < 1 % signal loss/month
    • Recommended Load Resistor 10 Ohm
    • Bias Voltage Not required
    • Repeatability < 2 % of signal
    • Output Linearity Linear


  • 4-series conforming standard size

    AWX CO-4 1000

    Mini 4-Series 3-pin configuration

    • Nominal Range 0 – 1000 ppm
    • Maximum Overload 2800 ppm
    • Expected Operation Life 2 years in air
    • Output Signal 50 +/- 15 nA/ppm
    • Inboard Filter To remove H2S and SO2
    • Resolution 2 ppm
    • Temperature Range - 20 °C to + 50 °C
    • Pressure Range Atmospheric +/-10 %
    • t90 Response Time < 25 sec
    • Relative Humidity Range 15 % to 90 % R.H. non-condensing
    • Typical Baseline Range (pure air, 20°C) - 5 ppm to +5 ppm
    • Maximum Zero Shift (+20°Cto +40 °C) 20 ppm equivalent
    • Expected Long Term Output
    • Drift < 2 % signal loss/month
    • Recommended Load Resistor 10 Ohm
    • Bias Voltage Not required
    • Repeatability < 2 % of signal
    • Output Linearity Linear


  • 4-series conforming standard size

    AWX CO-4C 500

    Mini 4-Series 2-pin configuration

    • Nominal Range 0 – 500 ppm
    • Maximum Overload 1000 ppm
    • Expected Operation Life 6 years in air
    • Output Signal 50 +/- 15 nA/ppm
    • Inboard Filter To remove H2S and SO2
    • Resolution 1 ppm
    • Temperature Range - 20 °C to + 50 °C
    • Pressure Range Atmospheric +/-10 %
    • t90 Response Time < 25 sec
    • Relative Humidity Range 15 % to 90 % R.H. non-condensing
    • Typical Baseline Range (pure air, 20°C) - 5 ppm to +5 ppm
    • Maximum Zero Shift (+20°Cto +40 °C) 10 ppm equivalent
    • Expected Long Term Output
    • Drift < 2 % signal loss/month
    • Recommended Load Resistor 10 Ohm
    • Bias Voltage Not required
    • Repeatability < 2 % of signal
    • Output Linearity Linear


  • 7-series conforming standard size

    AWX CO-7 1000

    Compact 7-Series

    • Nominal Range 0 – 1000 ppm
    • Maximum Overload 2000 ppm
    • Expected Operation Life 3 years in air
    • Output Signal 120 +/- 20 nA/ppm
    • Resolution 1.0 ppm
    • Temperature Range - 30 °C to + 30 °C
    • Pressure Range Atmospheric +/-10 %
    • t90 Response Time < 30 sec
    • Relative Humidity Range 15 % to 90 % R.H. non-condensing
    • Typical Baseline Range (pure air, 20°C) – +/- 7 ppm
    • Maximum Zero Shift (+20°C to +40 °C) 2.0 ppm
    • Expected Long Term Output Drift < 1 % signal loss/month
    • Recommended Load Resistor 10 Ohm
    • Bias Voltage Not required
    • Repeatability < 2 % of signal
    • Output Linearity Linear