AWX Cl2-3 20
Standard 3-Series
- Nominal Range 0 – 20 ppm
- Maximum Overload 200 ppm
- Expected Operation Life 2 years in air
- Output Signal -980 +/- 200 nA/ppm
- Resolution 0.05 ppm
- Temperature Range - 40 °C to + 50 °C
- Pressure Range Atmospheric +/-10 %
- t90 Response Time < 30 sec
- Relative Humidity Range 15 % to 90 % R.H. non-condensing
- Typical Baseline Range (pure air, 20°C) – 0.1 ppm to +0.2 ppm
- Maximum Zero Shift (+20°C to +40 °C) 0.5 ppm equivalent
- Expected Long Term Output Drift < 2 % signal loss/month
- Recommended Load Resistor 10 Ohm
- Bias Voltage Not required
- Repeatability < 2 % of signal
- Output Linearity Linear
-
AWX Cl2-4 20
Mini 4-Series
- Nominal Range 0 – 20 ppm
- Maximum Overload 200 ppm
- Expected Operation Life 2 years in air
- Output Signal -600 +/- 150 na/ppm
- Resolution 0.1 ppm
- Temperature Range - 40 °C to + 50 °C
- Pressure Range Atmospheric +/-10 %
- t90 Response Time < 50 sec
- Relative Humidity Range 15 % to 90 % R.H. non-condensing
- Typical Baseline Range (pure air, 20°C) -0.1 ppm to 0.2 ppm
- Maximum Zero Shift (+20°C to +40 °C) -0.2 ppm
- Expected Long Term Output Drift < 5 % signal loss/6 month
- Recommended Load Resistor 10 Ohm
- Bias Voltage Not required
- Repeatability < 2 % of signal
- Output Linearity Linear
-
AWX Cl2-7 20
Compact 7-Series
- Nominal Range 0 – 20 ppm
- Maximum Overload 200 ppm
- Expected Operation Life 2 years in air
- Output Signal -30 +/- 10 nA/ppm
- Resolution <0.05 ppm
- Temperature Range - 40 °C to + 50 °C
- Pressure Range Atmospheric +/-10 %
- t90 Response Time < 50 sec
- Relative Humidity Range 15 % to 90 % R.H. non-condensing
- Typical Baseline Range (pure air, 20°C) – 0.1 ppm to +0.2 ppm
- Maximum Zero Shift (+20°C to +40 °C) 0.5 ppm equivalent
- Expected Long Term Output Drift < 2 % signal loss/month
- Recommended Load Resistor 10 Ohm
- Bias Voltage Not required
- Repeatability < 2 % of signal
- Output Linearity Linear